The four main
liquid chromatography (LC) modes and normal and reversed phase operating
conditions and their respective mechanisms have been presented in previous posts.
After selection
of the LC mode, the selection of the proper column (stationary phase, particle
size, stability of the adsorbent) and operating parameters is the next step in
the development of a separation.
Based on the
outcome of the first trial chromatogram(s), a number of questions have to be
answered:
- Is there sufficient resolution of the desired sample components? If the components of interest are sufficiently resolved for qualitative, quantitative analysis or sample collection (in preparative LC chromatography) the chromatographer’s work is almost complete.
- Is the analysis obtained in a reasonable length of time? If there is sufficient resolution of the peaks then the flow rate of the mobile phase may be increased, or the column may be shortened so that the time required for elution will be shortened. Techniques for increasing speed, such as gradient elution can be used.
- Does the sample size permits detection of the smallest desired peak or a sufficient quantity of the desired component(s) for collection? If not, a longer column and/or higher capacity may be required.
A general
approach for column selection and optimization of operating parameters for
better separations will be presented in this post. The main steps are the
following:
Selection of the column and of the column packing
The selection of
the proper column is a crucial step in the separation process. The choice of
the packing is dictated by the requirements of the separation method. For
example, in analytical separations optimum resolution is required while sample
capacity is of less importance. In preparative separations column packing with
large sample capacity and optimum resolution is required at the expense of
speed.
An ideal column would
give excellent resolution for all sample components in the shortest time and
have the highest sample capacity required. However, since the three attributes
of chromatographic separations are resolution,
speed and column capacity the analyst must compromise. Generally, resolution
is optimized by sacrificing speed and sample size.
The basic column
parameters are given below:
Length: The column length
is an important design parameter. Doubling the length doubles the retention
time. However, resolution is only proportional to the square root of the column
length.
Diameter: Small diameter
columns are used to obtain better resolution and to improve speed. Typical HPLC
columns have a diameter 1-6 mm.
Solid support – Column packing: Column packing especially for HPLC falls into one of two main
categories – porous and pellicular porous.
Porous packing
materials are either spherical or irregularly shaped, have deep pores, and are
available in a wide range of particle sizes. As the diameter of the porous
particle decreases, the depth of the pore also decreases. The resulting shorter
diffusion path permits an increased rate of mass transfer so that column
efficiency and resolution increase.
The pellicular
porous packing materials provide an alternate way of decreasing diffusion
paths. They consist of a solid, non-porous core and a thin, porous outer shell
(Fig. 1). Rapid solute mass transfer occurs in the thin shell.
Fig. 1: Pellicular porous packing material. The solid
core is non-porous while the thin outer shell is porous.
|
Sample capacity of the column: The sample capacity of the column is directly related to the
quantity of available stationary phase. In LSC, the capacity is proportional to
the surface area of the adsorbent. In LLC, to the volume of liquid phase. In
ion-exchange chromatography, to the number of ion exchange sites.
Porous particles
have larger surface areas than non-porous ones. Columns of porous particles
have high sample capacity. A typical porous LSC adsorbent has a surface area of 200-400
m2/g. It can handle
samples in the range of mg/g of adsorbent without overloading.
Pellicular
porous packing has a low surface area and therefore a low sample capacity. A
typical pellicular porous support has a surface area of 7 m2/g.
The choice of
the column packing – porous or pellicular – is dependent on the sample capacity
requirement:
- If a sample contains six or more components and/or 0.5 mg or more of total sample, a porous packing should be used
- If a sample contains less than six components and /or less than 0.5 mg of total sample, a pellicular porous packing should be used.
In general, the
smaller the particle size the faster the rate of mass transfer and therefore
better efficiency. The minimum particle diameter which can be used is dependent
on the length and internal diameter of the column and the pressure capability
of the LC/HPLC pumping system.
In liquid
chromatography the particle size used is normally > 150 μm while in HPLC <40 μm ( in practice 3-8 μm).
References
L.R. Snyder. J.J. Kirkland, “Introduction to Modern Liquid
Chromatography”, 2nd edition, Wiley, 1979
A. Weston and P. Brown, “HPLC and CE Principles and Practice”,
Academic Press, 1997
C.F. Poole, S.K. Poole, “Chromatography Today”, Elsevier, New York,
1991
Nina Hadden et al., “Basic Liquid Chromatography”, Varian Aerograph, 1971
Nina Hadden et al., “Basic Liquid Chromatography”, Varian Aerograph, 1971