CALIBRATION

Calibration of an instrument in Analytical Chemistry is the operation that determines the functional relationship between measured values (signal intensities S at certain signal positions zi) and analytical quantities characterizing types of analytes qi and their amount (content, concentration) n. Calibration includes the selection of the model (its functional form), the estimation of the model parameters as well as the errors, and their validation.

During calibration, the instrument is “taught” using known values of calibrator standards what result it should provide. A calibration curve is constructed which is the graphical relationship between the known values, such as concentrations, of a series of calibration standards and their instrument response. A calibration software is also can be used for automating calibration processes plus managing and reporting results for various measurement disciplines including electrical, temperature, pressure, flow and RF.

Proper calibration is important because ensures that an instrument (analytical device) remains within validated performance limits to accurately report results. Calibration is one of the primary processes used to maintain instrument accuracy.


References

  1. K. Danzer, L.A. Currie, Pure & Appl. Chem., Vol. 70, 4, 993-1014 (1998
  2. D. Harvey, “Modern Analytical Chemistry”, McGraw-Hill Companies Inc., 2000
  3. D.A. Skoog, F.J. Holler, T.A. Nieman, “Principles of Instrumental Analysis”. Saunders College Publishing: Philadelphia, 1998.

Author: | Publisher: Chemistry Net

Quantitative Analysis by Atomic Absorption

Quantitative Analysis by Atomic Absorption

 

Quantitative Analysis by Atomic Absorption

 

The capability of an atom to absorb very specific wavelengths of light is utilized in atomic absorption spectrometry.

Light of a specific wavelength and of initial intensity Io is focused on the flame cell containing ground state atoms. The initial light intensity is decreased by an amount determined by the atom concentration in the flame cell. The light is then directed to the detector where the reduced intensity, It, is measured. The amount of light absorbed is determined by comparing It  to Io and according to Beer’s law:

 

A = log Io / It = α * c * d         (1)

 

Absorbance A is the most convenient term for characterizing light absorption in absorption spectrophotometry, as this quantity follows a linear relationship with concentration c.

Equation (1) can be used for quantitative analysis by atomic absorption spectrometry AAS. A calibration curve is used to determine the unknown concentration of an element – i.e. nickel – in a solution. The instrument is calibrated using several solutions of known concentrations of the element under examination. The calibration curve shows the concentration of the element in solution against the amount of radiation absorbed (Fig. I.1).

The sample solution is fed into the instrument and the unknown concentration of the element  – i.e. nickel – is then displayed on the calibration curve (Fig. I.1).

For example for the Ni solution above an absorbance of 0.37 was obtained that corresponds to a concentration of 12 mg/l.

Over the region where the Beer’s law relationship is observed, the calibration yields a straight line. As the concentration and absorbance increase, nonideal behavior in the absorption process can cause a deviation from linearity as shown in Fig. I.1. There are several reasons for this nonideal behavior  such as nonhomogeneities of temperature and space in the absorbing cell, line broadening, absorption at nearby lines and stray light.

As shown above, after such a calibration is established (Fig. I.1) the absorbance of solutions of unknown concentrations may be measured and the corresponding concentrations can be determined from the calibration curve.

The instrument performance for an element can be monitored by the following parameters:

  • Characteristic concentration for the element
  • Detection limit

The characteristic concentration for an element (called “sensitivity”) is a convention and is defined as the concentration of analyte giving an absorbance of 0.00436 (corresponding to a percent transmittance of 99%). Usually the wavelength providing the best sensitivity is used, although a less sensitive wavelength may be more appropriate for a high concentration of analyte. A less sensitive wavelength also may be appropriate when significant interferences occur at the most sensitive wavelength.

 

Characteristic Conc. (mg/l) = Conc. of Standard (mg/l) * 0.0044  /  measured absorbance

 

There are several practical reasons for wanting to know the value of the characteristic concentration for an element. For example, knowing the expected characteristic concentration of an element allows an operator to determine if all instrumental conditions are optimized and if the instrument is performing according to specifications. This is accomplished by simply measuring the absorbance of a known concentration of the element and comparing the results to the expected value.

Even though the magnitude of the absorbance signal can be predicted from the value given for characteristic concentration, no information is given on how small of an absorbance signal can be measured.

The smallest measurable concentration of an element – the detection limit of the element -  will be determined by the magnitude of the absorbance observed for the element and the stability of the absorbance signal.

The detection limit (according to IUPAC) is the smallest concentration or absolute amount of analyte that has a signal significantly larger than the signal arising from a reagent blank.

Mathematically, the analyte’s signal at the detection limit (sDL) is given by:

 

sDL = sreag + 3 * σreag

where sreag is the signal for a reagent blank, sreag is the known standard deviation for the reagent blank’s signal.

Other approaches for defining the detection limit have also been developed. In atomic absorption spectrometry usually the detection limit is determined for a certain element by analyzing a diluted solution of this element and recording the corresponding absorbances. The experiment is repeated for 10 times. The 3σ of the recorded absorbance signal can be considered as the detection limit for the specific element under the experimental conditions used – wavelength, type of flame, instrument.

For example, let us suppose that the detection limit for Cu has to be determined by AAS under certain experimental conditions. A 0.1 ppm Cu solution is analyzed for 10 times by AAS at a wavelength of 324.8 nm and the corresponding absorbance values are recorded:


Experiment #
Absorbace
1
0.006
2
0.005
3
0.007
4
0.007
5
0.006
6
0.007
7
0.005
8
0.004
9
0.005
10
0.004
Average
0.0056
σ
0.0012
3σ
0.0036

 

Therefore, the detection limit for Cu (Cu signal at the detection limit) under the above conditions is 0.0036 and this absorbance corresponds to a Cu solution concentration approximately at:

0.1 ppm * (0.0036/0.0056) = 0.064 ppm (assuming we are working on the linear region of the calibration curve).


Relevant Posts

Atomic Absorption Spectrometry (AAS)

Advantages and Disadvantages of Atomic Absorption Spectrometry and Graphite Furnace

 


References

  1. R. Ferrus, ; M.R. Egea, Anal. Chim. Acta, 287, 119–145 (1994)
  2. J.A. Glaser, D.L. Foerst, et al. Environ. Sci.Technol., 15, 1426–1435 (1981)
  3. P.W.J. Boumans, Anal. Chem., 66, 459A–467A (1994)

What is calibration? Calibrated instruments|Analytical Chemistry



Why calibration is such a common and important step in analytical methods? How is defined?

Calibration is the process of assigning a value, usually in concentration units, to an instrument response.


For example, you might calibrate the response of an analytical device – such as a spectrophotometer - by analyzing different “known” concentrations of an analyte (i.e. solution of a metal ion such as Zn+2) and establish the instrument’s response per unit concentration of the metal (Fig. 1). Then, an unknown concentration of a test sample of the same analyte can be estimated from the graph by extrapolating the observed absorbance (y-axis) over to the analyte concentration (x-axis) (yellow line in Fig. 1).
In the graph shown in Fig. 1, the calibration function is established based on points in Table I.1:
                     

Table I.1: Instrument Response for different analyte A  concentrations

 Analyte Concentration (μg/l)
Instrument Response (Absorbance)
0
0
2
0.120
4
0.240
6
0.360
8
0.480
10
0.600


The line in Fig. 1 is called calibration line or in general calibration curve.
Calibration line of an instrument's absorbance vs. Analyte's Concentration
Fig.1: A calibration line of an instrument's absorbance A vs. Analyte's Concentration


Calibration curve is the graphical relationship between the known values, such as concentrations, of a series of calibration standards and their instrument response. 
 


What is the definition of calibration?

A definition of calibration according to IUPAC1 is as follows:

Calibration in Analytical Chemistry is the operation that determines the functional relationship between measured values (signal intensities S at certain signal positions zi) and analytical quantities characterizing types of analytes qi and their amount (content, concentration) n. Calibration includes the selection of the model (its functional form), the estimation of the model parameters as well as the errors, and their validation.

How an instrument – analytical device is calibrated?

As mentioned above, normally a calibration graph of a standard A response versus different concentrations of the standard A is constructed.
The analyte –standard A - response  may be: i) absorbance in U.V spectrometers  and atomic absortption spectrophotometers  ii) peak area (or height) or peak area (height) ratios to an internal standard in chromatographic assays.
The relationship may be linear (a straight line) or curved.
In the past – before ready access to personal computers and graph constructing softwares – calibration curves were drawn on graph paper and the concentrations interpolated manually. At that time having a straight line made the task considerably easier and scientists tried to work within a “linear” region of the curve even for techniques that are not linear (fluorescence, ECD).
Even today the ideal case is a linear response that passes directly through the origin comparing to the non-linear calibrations since the equation of a line is exactly known and therefore errors are minimized .

In case that the relationship is linear will be of the form:

(Smeas)A = k * nA + Sreag      (1)

or


(Smeas)A = k * CA+ Sreag      (2)

Notice that the above equations are of the form: y = m * x + b  (equation of a straight line). That means that k is the slope m of the straight line and Sreag is the intercept b of the line – the point where the line crosses the y axis.
Where (Smeas)A is the measured signal of substance A, k is a proportionality constant, nA or CA are the moles or concentration respectively of analyte A and Sreag is the signal due to the reagents (constant error)2.

In the special case that Sreag = 0 then the straight line passes through zero as in Figure 1.
The equation of the calibration line –straight line – in Fig. 1 is given by:

(Smeas)A = k * CA+ Sreag   or  

      y = m * x + b             (3)
 
The slope of the line = m = k = Rise/Run = 0.6/10 = 0.06 (determined directly from the graph in Fig.1). The intercept b = 0

Therefore (3) becomes      y = 0.06 * x
If you plug in values of x (analyte’s concentration) from Table I.1 in equation (3) you will get the corresponding values of y (instrument response) shown in the Table.

Normally though, the points from which the calibration function is determined do not fit so well as presented above.
Let us consider that the instrument’s response to analyte’s A concentrations are those given in Table I.2. Notice that for the standard solution with no analyte present the instrument response was 0.0123 – this standard is called a blank.

A blank is a specimen that is intended to contain none of the analytes of interest and which is subjected to the usual analytical or measurement process to establish a zero baseline or background value.

Table I.2: Instrument Response for different analyte A  concentrations

 Analyte Concentration (μg/l)
Instrument Response (Absorbance)
0
.0123
2
.1244
4
.2507
6
.3820
8
.4886
10
.5995
 
In a case like this it is always good practice to plot data before carrying out any statistical analysis (regression). Plot just a scatter plot (Figure 2 – a scatter plot shows only the points and not the best fitted line) using Excel or any statistical software (SPSS, Sigmaplot…). 


Are there any outliers in the calibration function?

You will note from the graph in Fig. 2 that there is some scattering around the best-fitted line and the line crosses the y axis not at zero as before. This is the normal situation for experimental data.
Since the best-fitted line is affected considerably by outliers – introduce significant errors - we must check the graph for the presence of possible outliers.

 Absorbance of analyte A vs. Analyte A Concentration  (linear best fitted line)

Fig. 2: Absorbance of analyte A vs. Analyte A Concentration (linear best fitted line)



An outlier is a result which is significantly different from the rest of the data set. In the case of calibration, an outlier would appear as a point which is well removed from the other calibration points.

There is no outlier in Fig 2, just some scattering that is normal for experimental data.

The effects outliers can have on the best fitted calibration line are leverage and bias.

For example an outlier at the extremes of the calibration range can change the position of the calibration line by tilting it upwards or downwards (Table I.2a, Fig 2a, ). The instrument response at analyte concentration 10 μg/l is such an outlier. 


Table I.2a: Instrument Response for different analyte A  concentrations

 Analyte Concentration (μg/l)
Instrument Response (Absorbance)
0
.0123
2
.1244
4
.2507
6
.3820
8
.4886
10
.9995

This outlier is said to have a high degree of leverage. Leverage affects both the slope and the intercept of the calibration line and introduces errors.

An outlier in the middle of the calibration range will shift the best fitted line up or down. The outlier in this case introduces a bias into the position of the line – an additive error. The slope of the line will not be affected though but the intercept will be wrong.

Absorbance of analyte vs. analyte concentration. The outlier point (white point) was excluded when the best-fitted line was drawn (black line). If it will be included the best-fitted line is the red line

Fig. 2a: Absorbance of analyte A vs. Analyte A Concentration (linear best fitted line). The outlier point (white point at (10, 0.9995)) that corresponds to analyte concentration = 10 μg/l was excluded when the best-fitted line was drawn (black line). When this outlier point is included in the calculations for the best fitted line the line changes position and the slope and intercept change considerably (red line).

 
How do you correct for outliers in the calibration graph?

Use one of the following methods, preferably the last three:

1)   Remove with caution points that are significantly different from the rest of the data. For example the outlier point in Fig. 2a can be excluded and the best fitted line should be drawn through the rest of the points.
2)    Repeat the calibration procedure and get the average of 5 or more replicate measurements at each concentration level (concentration levels ≥ 3). Then plot the averages of these replicate measurements at each concentration level.
3)      Use the method of standard additions.
4)      Use the bracketing method.
 Then draw the best fitted calibration line checking first if the calibration is linear.